7-18 Keysight N5247B Service Guide
Repair and Replacement Procedures
Removing and Replacing the A4–A17 Boards
7-
Removing and Replacing the A4–A17 Boards
Tools Required
— T-10 TORX driver (set to 9 in-lb)
— T-20 TORX driver (set to 21 in-lb)
— 5/16 inch open-end torque wrench (set to 10 in-lb)
— 9 mm socket or open-end wrench (set to 21 in-lb)
— ESD grounding wrist strap
Removal Procedure
If you have a serial number prefix
<
6021, refer to
All Options, S/N Prefixes <6021”
Figure 6-6, “Top Assemblies, All Options,
for this procedure. If you have a serial number prefix
≥
6021, refer to
Figure 6-8, “Top Assemblies, All Options, S/N Prefixes
Regardless of your serial number prefix, if you have had your instrument
upgraded to a version 7 synthesizer assembly, refer to
Assemblies, All Options, S/N Prefixes
1.
Disconnect the power cord.
2.
Remove the outer and inner covers. Refer to
.
3.
Instructions for A5 and A10 source boards (Refer to
for cable locations):
a.
A10 source board only: on the bottom side of the analyzer,
disconnect the cables from the A12 and A13 boards (item
).
b.
A5 source board only: on the top side of the analyzer, disconnect all
visible semirigid cables from the A4, A7, and A8 boards.
c.
On the top side of the analyzer, remove two screws (item
), one at
each end of the board, from the source board to be removed.
d.
Lift the two extractors (item
), one at each end of the board.
Adjust the slack in the gray flexible cable as needed to move it out
of the way, and lift the board out of the chassis.
If you have a PNA with a serial number prefix
<
6021 and have not had a
2S7 or 4S7 version 7 synthesizer
assembly
upgrade, then you have A4,
A14, A15, and A17 boards.
If you have a PNA with a serial number prefix
≥
6021 or have had a 2S7 or
4S7 version 7 synthesizer assembly upgrade, then you have an A15 direct
digital synthesizer (DDS) assembly instead of the A4, A14, A15, and A17
boards.
Refer to
“Top Assemblies and Cables, All Options, Serial Number Prefixes
Summary of Contents for N5247B
Page 10: ...10 Keysight N5247B Service Guide Contents ...
Page 96: ...3 66 Keysight N5247B Service Guide Tests and Adjustments Adjustments 3 ...
Page 157: ...Keysight N5247B Service Guide 4 61 Troubleshooting This page left intentionally blank ...
Page 268: ...6 30 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 270: ...6 32 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 373: ...Keysight N5247B Service Guide 6 135 Replaceable Parts Replaceable Parts Listings ...
Page 426: ...6 188 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 565: ...Installation Note Xxxxx xxxxx 3 ...