Keysight N5247B Service Guide 4-15
Troubleshooting
Front Panel Troubleshooting
Front Panel Troubleshooting
The front panel assembly consists of the A1 front panel interface board, the
keypad, the A2 USB board, and the A3 display assembly. The following tests
verify the operation of the front panel assembly when the analyzer is in the
measurement mode. If the instrument fails to power up correctly, or it is
difficult to verify due to a faulty display, refer to
.
Refer to the following sections to verify the operation of the noted assemblies.
—
“Front Panel Keypad and RPG Test”
on this page
—
“A3 Display Test” on page 4-18
—
“Checking the A2 USB Board” on page 4-19
—
“A1 Front Panel Interface Board” on page 4-20
If all assemblies are working correctly, continue troubleshooting with
Panel Troubleshooting” on page 4-22
.
Front Panel Keypad and RPG Test
Test the front panel keypad by running the front panel test. To run the front
panel test, perform the following:
Press UTILITY
, then
, then
, then
.
A
Front Panel Key Test Utility
dialog box will be displayed, as shown in
.
System
Service
Diagnostics
Front Panel Test
Summary of Contents for N5247B
Page 10: ...10 Keysight N5247B Service Guide Contents ...
Page 96: ...3 66 Keysight N5247B Service Guide Tests and Adjustments Adjustments 3 ...
Page 157: ...Keysight N5247B Service Guide 4 61 Troubleshooting This page left intentionally blank ...
Page 268: ...6 30 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 270: ...6 32 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 373: ...Keysight N5247B Service Guide 6 135 Replaceable Parts Replaceable Parts Listings ...
Page 426: ...6 188 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 565: ...Installation Note Xxxxx xxxxx 3 ...