3-38 Keysight N5247B Service Guide
Tests and Adjustments
Performance Tests
3-
13.
The magnitude and phase measurements using the receiver under test are
read: (Pd).
14.
The compression for each point is calculated as (Pa–Pb) – (Pd–Pc).
If the Analyzer Fails this Test:
— Run the Receiver Characterization adjustment, and repeat this test.
— If the analyzer still fails this test, replace the A23 mixer brick for a Port 1 or
Port 2 failure or the A24 mixer brick for a Port 3 or Port 4 failure, then repeat
“Removing and Replacing the A27 and A28 Mixer Bricks”
Noise Floor Test
Function of the Test: To measure the absolute power level of the noise floor for
the analyzer’s receivers.
Specification Tested: Test Port Input–Test Port Noise Floor
Equipment Used:
— Power meter
— Power sensors
— Test cable
— Calibration kit
Description of the Test:
This test uses the source signal out of one analyzer test port as part of the
noise floor measurement on another test port. Port 2 is the source port when
measuring the noise floor of Port 1. Port 1 is the source port when measuring
the noise floor of Ports 2, 3, and 4.
1.
The analyzer is Preset.
2.
The analyzer is set up for a CW transmission measurement between the
source port and the test port to be measured. The analyzer is set to an IF
bandwidth of 1 kHz and 801 points per sweep.
3.
A test cable is connected to the source port.
4.
A power sensor is connected to the end of the test cable with any
necessary adapters.
5.
For hundreds of frequencies across the analyzer’s full range, a source
power calibration is performed to ensure a flat power response at the end
of the cable at the Preset power level.
6.
The power sensor is disconnected and the cable is connected to the port
to be tested.
Summary of Contents for N5247B
Page 10: ...10 Keysight N5247B Service Guide Contents ...
Page 96: ...3 66 Keysight N5247B Service Guide Tests and Adjustments Adjustments 3 ...
Page 157: ...Keysight N5247B Service Guide 4 61 Troubleshooting This page left intentionally blank ...
Page 268: ...6 30 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 270: ...6 32 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 373: ...Keysight N5247B Service Guide 6 135 Replaceable Parts Replaceable Parts Listings ...
Page 426: ...6 188 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 565: ...Installation Note Xxxxx xxxxx 3 ...