7-12 Keysight N5247B Service Guide
Repair and Replacement Procedures
Removing and Replacing the A1–A3 and Other Front Panel Subassemblies
7-
Removing the Speaker Assembly
1.
Disconnect the speaker cable from the interface board.
2.
Remove the speaker assembly and speaker foam.
Removing the A3 Display Assembly and the Touchscreen
1.
Disconnect the following cables from the A1 front panel interface board:
tape (item [14]) from connector and the LCD display to interface board
cable (item [18]), LED driver board (backlight converter board) to interface
board (item [11]), touch screen controller board to interface board cable
(item [13]), power switch board to interface board cable (item [12]), and
the speaker assembly cable.
2.
Remove eleven outer screws (item
) from the A3 display assembly’s
mounting plate and remove the A3 display assembly from the front panel
assembly.
3.
Raise the retaining clamps on the LCD display flat flex cable’s two
connectors (item [15]).
4.
Disconnect the flat flex cable from the touchscreen (item [15]).
5.
The LED driver board cable (item 22) can be removed from the connection
on the LCD Display.
6.
The LCD display cable (item [18]) can be removed by removing the three
screws (item [17]) that attach it to the LCD display mounting bracket.
7.
Remove the 3 screws on the slotted hole side of the mounting plate (item
[20]).
8.
Remove the 3 screws on the round hole side of the mounting plate (item
[20]).
9.
With a T-6 TORX (set to 3-in-lb), remove the 2 screws (item [23]) on the
side bracket that has the LCD cable (item [18]) attached to the LCD
display.
When handling the speaker avoid touching the top of the speaker. Hold
only the sides of the speaker when re-attaching the new foam.
Avoid covering the Mounting Plate screw.
During re-assembly of the side brackets to the LCD display, it is very
important to avoid over-tightening.
Summary of Contents for N5247B
Page 10: ...10 Keysight N5247B Service Guide Contents ...
Page 96: ...3 66 Keysight N5247B Service Guide Tests and Adjustments Adjustments 3 ...
Page 157: ...Keysight N5247B Service Guide 4 61 Troubleshooting This page left intentionally blank ...
Page 268: ...6 30 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 270: ...6 32 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 373: ...Keysight N5247B Service Guide 6 135 Replaceable Parts Replaceable Parts Listings ...
Page 426: ...6 188 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 565: ...Installation Note Xxxxx xxxxx 3 ...