Keysight N5247B Service Guide 3-53
Tests and Adjustments
DC Continuity Test the LFE Board and Test Ports (Available With Low Frequency
Extension (LFE) Installed Option 205/425)
DC Continuity Test the LFE Board and Test Ports (Available With
Low Frequency Extension (LFE) Installed Option 205/425)
The DC continuity test verifies that the LFE board is installed correctly and
does not have any opens or shorts in the DC path.
1.
Using a DVM, connect one test probe to the center conductor of the RF
port 1 on the front panel.
2.
Connect the other test probe to the port 1 bias input (
BIAS 1 IN
) on the
rear panel.
3.
Verify the DVM measures
<
10
Ω
.
4.
Repeat these steps for each of the other test ports.
If the DVM value is 0
Ω
or
>
10
Ω
, then something is incorrectly installed or
there is an open or short somewhere in the LFE board/cable path:
—
Verify the gray DC bias cables installed in
Cables, All Options, Serial Number Prefixes <6021” on page 6-18
and the blue RF flexible cables installed in
in the “Bottom
Assemblies” section—for your LFE option—are connected correctly
and not open or shorted.
Summary of Contents for N5247B
Page 10: ...10 Keysight N5247B Service Guide Contents ...
Page 96: ...3 66 Keysight N5247B Service Guide Tests and Adjustments Adjustments 3 ...
Page 157: ...Keysight N5247B Service Guide 4 61 Troubleshooting This page left intentionally blank ...
Page 268: ...6 30 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 270: ...6 32 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 373: ...Keysight N5247B Service Guide 6 135 Replaceable Parts Replaceable Parts Listings ...
Page 426: ...6 188 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 565: ...Installation Note Xxxxx xxxxx 3 ...