3-22 Keysight N5247B Service Guide
Tests and Adjustments
System Verification
3-
System Verification
System verification is used to verify system-level, error-corrected uncertainty
limits for network analyzer measurements. The verification procedure is
automated and is contained in the firmware of the analyzer.
The device data provided with the verification kit has a traceable path to a
national standard. The difference between the supplied traceable data and the
measured data must fall within the total uncertainty limits at all frequencies for
the system verification to pass.
The total measurement uncertainty limits for the system verification are the
sum of the factory measurement uncertainties for the verification devices and
the uncertainties associated with the system being verified. You can determine
your system measurement uncertainty limits by referring to the analyzer
embedded on-line help.
What the System Verification Verifies
The system verification procedure verifies proper operation of the:
— network analyzer
— calibration kit
— test port cables
together as a “system”. It DOES NOT verify that any of these components pass
their specifications independently. The user is responsible for independently
calibrating and verifying the proper operation of the calibration kit and test
port cables prior to performing the system verification.
Measurement Uncertainty
Measurement uncertainty is defined as the sum of:
— the residual systematic (repeatable) errors, and
— the random (non-repeatable) errors
in the measurement system after calibration.
The systematic errors are:
Passing this system verification does not guarantee that the analyzer
meets all of its performance specifications. However, it does show that the
network analyzer being verified measures the same devices with the same
results as a factory system which has had all of its specifications verified
and its total measurement uncertainty minimized.
Additional equipment or accessories used with the above system are not
verified by system verification.
Summary of Contents for N5247B
Page 10: ...10 Keysight N5247B Service Guide Contents ...
Page 96: ...3 66 Keysight N5247B Service Guide Tests and Adjustments Adjustments 3 ...
Page 157: ...Keysight N5247B Service Guide 4 61 Troubleshooting This page left intentionally blank ...
Page 268: ...6 30 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 270: ...6 32 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 373: ...Keysight N5247B Service Guide 6 135 Replaceable Parts Replaceable Parts Listings ...
Page 426: ...6 188 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 565: ...Installation Note Xxxxx xxxxx 3 ...