Theory of Operation
Signal Separation Group Operation
5-
Signal Separation Group Operation
The signal separation group divides the source incident signals into a reference
path and a test path. Refer to
and
— The reference signals are transmitted to the receiver group as the R1, R2,
R3, and R4 inputs for 4-port models or the R1 and R2 inputs for 2-port
models.
— The test signals are transmitted through—and reflected from—the device
under test (DUT) and then transmitted to the receiver group as the A, B, C,
and D inputs for 4-port models or the A and B inputs for 2-port models.
— Control lines to this group are routed from the A23 test set motherboard.
In this section, the following assemblies are described:
—
—
— Front Panel Jumpers—
—
A38–A41 50-dB Source Step Attenuators and A42–A45 Bias Tees
—
A38–A41 50-dB Source Step Attenuators and A71–74 Bias Tee Combiners
—
A50–A53 Mechanical Switches and A54 Combiner (Optional)
Configurable Test Set
The configurable test set is included in the standard analyzer and allows you to
measure devices with higher power and higher dynamic range limits than an
analyzer without the configurable test set. On 4-port models, twelve signal
paths, routed through front panel SMA jumpers, comprise the configurable test
set. On 2-port models, there are six signal paths routed through front panels
jumpers in the configurable test set.
As shown in
and
jumpers are installed between the components listed below. Ports 3 and 4
apply only to 4-port models.
— the A29 port 1 reference coupler and the A27 mixer brick receiver R1
— the A30 port 3 reference coupler and the A28 mixer brick receiver R3
— the A31 port 4 reference coupler and the A28 mixer brick receiver R4
— the A32 port 2 reference coupler and the A27 mixer brick receiver R2
— the A29 port 1 reference coupler and the A33 test port 1 coupler
— the A30 port 3 reference coupler and the A34 test port 3 coupler
Summary of Contents for N5247B
Page 10: ...10 Keysight N5247B Service Guide Contents ...
Page 96: ...3 66 Keysight N5247B Service Guide Tests and Adjustments Adjustments 3 ...
Page 157: ...Keysight N5247B Service Guide 4 61 Troubleshooting This page left intentionally blank ...
Page 268: ...6 30 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 270: ...6 32 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 373: ...Keysight N5247B Service Guide 6 135 Replaceable Parts Replaceable Parts Listings ...
Page 426: ...6 188 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 565: ...Installation Note Xxxxx xxxxx 3 ...