Keysight N5247B Service Guide 4-9
Troubleshooting
Power Up Troubleshooting
Figure 4-2
A19 Midplane Board Power Supply LED Indicators and Measurement Points
The instrument contains potentially hazardous voltages. Refer to the
safety symbols provided on the instrument and in
before operating the unit with the cover
removed. Make sure that the safety instructions are strictly followed.
Failure to do so can result in personal injury or loss of life.
Do not operate the analyzer with the outer cover removed for more than
30 minutes, as this could cause the analyzer to overheat which could
result in costly damage.
any one individual voltage supply from the A20 power supply develops an
over-voltage or over-current problem, all supplies are affected. The cause
of the over-voltage or over-current condition can be the A20 power supply
itself, or any assembly to which the A20 power supply provides voltage. To
isolate the cause, continue to the assembly removal process as described
Summary of Contents for N5247B
Page 10: ...10 Keysight N5247B Service Guide Contents ...
Page 96: ...3 66 Keysight N5247B Service Guide Tests and Adjustments Adjustments 3 ...
Page 157: ...Keysight N5247B Service Guide 4 61 Troubleshooting This page left intentionally blank ...
Page 268: ...6 30 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 270: ...6 32 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 373: ...Keysight N5247B Service Guide 6 135 Replaceable Parts Replaceable Parts Listings ...
Page 426: ...6 188 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 565: ...Installation Note Xxxxx xxxxx 3 ...