Keysight N5247B Service Guide 4-55
Troubleshooting
Measurement System Troubleshooting
Checking the A24 IF Multiplexer Board
1.
Locate each of the flexible RF cables at the output receivers of the IF
multiplexer board:
—
2-port models
—
Receiver A; W69
—
Receiver R1; W72
—
Receiver R2; W73
—
Receiver B; W70
—
4-port models
—
Receiver A; W69
—
Receiver B; W70
—
Receiver C; W72
—
Receiver D; W73
—
Receiver R; W71
2.
Disconnect the appropriate flexible RF cable from the output receiver to be
tested on the A24 IF multiplexer board.
3.
Connect the spectrum analyzer to the open connector.
4.
Refer to
. The measured signal on the
spectrum analyzer should be at 7.61 MHz (analyzers with DSP version
4.0), or 7.44 MHz (analyzers with DSP version 5.0), and 1 GHz.
5.
If the measured signal is present, replace the A16 SPAM board. Refer to
“Removing and Replacing the A4–A17 Boards” on page 7-18
.
6.
If the measured signal is not present, replace the A24 IF multiplexer board.
Refer to
“Removing and Replacing the A24 IF Multiplexer Board” on
.
Summary of Contents for N5247B
Page 10: ...10 Keysight N5247B Service Guide Contents ...
Page 96: ...3 66 Keysight N5247B Service Guide Tests and Adjustments Adjustments 3 ...
Page 157: ...Keysight N5247B Service Guide 4 61 Troubleshooting This page left intentionally blank ...
Page 268: ...6 30 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 270: ...6 32 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 373: ...Keysight N5247B Service Guide 6 135 Replaceable Parts Replaceable Parts Listings ...
Page 426: ...6 188 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 565: ...Installation Note Xxxxx xxxxx 3 ...