4-36 Keysight N5247B Service Guide
Troubleshooting
Measurement System Troubleshooting
4-
Equipment Setup
1.
Before starting these checks, zero and calibrate the power meter. (See the
power meter user’s guide for instructions on setting the calibration factor.)
2.
If the
Receiver Display
) is not on the analyzer
screen, perform the following: Press UTILITY
, then
, then
, then
.
3.
Set the sweep speed for a 10 second sweep: Press STIMULUS
,
then
. Set the time to 10.000 seconds in the
Sweep
Time
box.
To isolate a broadband RF signal generation failure, check the test port
output power:
1.
Note the power reading displayed on the power meter; it should be the
preset power level +/- 1 dB.
2.
Connect the power sensor, in turn, to Ports 2, 3, and 4 and set trace to
measure S22, S33, and S44 respectively. Note the power reading
displayed on the power meter.
— If the power level is low or high on all test ports, the problem is LO signal
related. Continue with
“Checking the A14 50 MHz Reference Outputs”
— If the power level is low or high on only one of the test ports, the problem is
either source group or in the signal separation group. Continue with
“Checking the A14 50 MHz Reference Outputs”
to check the source group.
Equipment Type
Model or
Part Number
Alternate Model or
Part Number
Power meter
E4418B/E4419B
E4418A/E4419A
Power sensor, 2.4 mm
E8487A
None
Adapter, 2.4 mm (f) to 2.4 mm (f)
11900B
85056-60007
System
Service
Utilities
Receiver Display
Sweep
Sweep Time
Summary of Contents for N5247B
Page 10: ...10 Keysight N5247B Service Guide Contents ...
Page 96: ...3 66 Keysight N5247B Service Guide Tests and Adjustments Adjustments 3 ...
Page 157: ...Keysight N5247B Service Guide 4 61 Troubleshooting This page left intentionally blank ...
Page 268: ...6 30 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 270: ...6 32 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 373: ...Keysight N5247B Service Guide 6 135 Replaceable Parts Replaceable Parts Listings ...
Page 426: ...6 188 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 565: ...Installation Note Xxxxx xxxxx 3 ...