4-4 Keysight N5247B Service Guide
Troubleshooting
Operating the Analyzer With Covers Removed
4-
Operating the Analyzer With Covers Removed
Protect Against Electrostatic Discharge (ESD)
This is important. If not properly protected against, electrostatic discharge can
seriously damage your analyzer, resulting in costly repair.
Assembly Replacement Sequence
After identifying the problem requiring an assembly to be replaced, follow
these steps:
Step 1.
Order a replacement assembly. Refer to
Chapter 6, “Replaceable Parts.”
Step 2.
Replace the faulty assembly and determine what adjustments are necessary.
Chapter 7, “Repair and Replacement Procedures.”
Step 3.
Perform the necessary adjustments. Refer to
Step 4.
Perform the necessary performance tests. Refer to
Do not operate the analyzer with the outer cover removed for more than
30 minutes, as this could cause the analyzer to overheat which could
result in costly damage.
To reduce the chance of electrostatic discharge, follow all of the
“Electrostatic Discharge Protection” on
, for all of the procedures in this chapter.
Summary of Contents for N5247B
Page 10: ...10 Keysight N5247B Service Guide Contents ...
Page 96: ...3 66 Keysight N5247B Service Guide Tests and Adjustments Adjustments 3 ...
Page 157: ...Keysight N5247B Service Guide 4 61 Troubleshooting This page left intentionally blank ...
Page 268: ...6 30 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 270: ...6 32 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 373: ...Keysight N5247B Service Guide 6 135 Replaceable Parts Replaceable Parts Listings ...
Page 426: ...6 188 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 565: ...Installation Note Xxxxx xxxxx 3 ...