4-38 Keysight N5247B Service Guide
Troubleshooting
Measurement System Troubleshooting
4-
c.
Verify that all cable connections to the A15 DDS assembly are
connected properly.
d.
Repeat steps 1 through.3.
e.
If the instrument still has a similar “Slug” failure then the DDS is
faulty and needs to be replaced.
5.
If the A15 DDS assembly is faulty, contact Keysight Technologies for help.
Refer to
“Contacting Keysight” on page 2-7
.
Figure 4-10
Example of Power On Self Test Error