Series 2600B System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2600BS-901-01 Rev. B / May 2013
7-87
Example
print(eventlog.enable)
eventlog.enable = eventlog.DISABLE
print(eventlog.enable)
Displays the present status of the Series
2600B event log.
Output:
1.00
0.00
Also see
eventlog.all()
(on page 7-84)
eventlog.clear()
(on page 7-85)
eventlog.count
(on page 7-86)
eventlog.next()
(on page 7-87)
eventlog.overwritemethod
eventlog.next()
This function returns the oldest message from the event log and removes it from the log.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
logString
= eventlog.next()
logString
The next log entry
Details
Returns the next entry from the event log and removes it from the log.
If there are no entries in the event log, returns the value
nil
.
Example 1
print(eventlog.next())
Get the oldest message in the event log and remove that entry from the log.
Output:
17:28:22.085 10 Oct 2009, LAN2, 192.168.1.102, LXI, 0, 1192037134, <no time>, 0,
0x0
Example 2
print(eventlog.next())
If you send this command when there is nothing in the event log, you will get the following output:
nil
Also see
eventlog.all()
(on page 7-84)
eventlog.clear()
(on page 7-85)
eventlog.count
(on page 7-86)
eventlog.enable
(on page 7-86)
eventlog.overwritemethod
Summary of Contents for System SourceMeter 2601B