Series 2600B System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2600BS-901-01 Rev. B / May 2013
7-55
Example
digio.trigger[4].pulsewidth = 20e-6
Sets the pulse width for trigger line 4 to
20
μs.
Also see
digio.trigger[N].assert()
(on page 7-51)
digio.trigger[N].reset()
(on page 7-55)
digio.trigger[N].release()
(on page 7-55)
digio.trigger[N].release()
This function releases an indefinite length or latched trigger. This function is not available on the Models
2604B/2614B/2634B.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
digio.trigger[
N
].release()
N
Digital I/O trigger line (1 to 14)
Details
Releases a trigger that was asserted with an indefinite pulse width time. It also releases a trigger that was
latched in response to receiving a synchronous mode trigger. Only the specified trigger line is affected.
Example
digio.trigger[4].release()
Releases digital I/O trigger line 4.
Also see
digio.trigger[N].assert()
(on page 7-51)
digio.trigger[N].pulsewidth
digio.trigger[N].reset()
This function resets trigger values to their factory defaults. This function is not available on the Models
2604B/2614B/2634B.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
digio.trigger[
N
].reset()
N
Digital I/O trigger line (1 to 14)