Series 2600B System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2600BS-901-01 Rev. B / May 2013
7-279
Details
These attributes are used to read or write to the operation status LAN summary registers. The binary equivalent
of the value indicates which register bits are set. In the binary equivalent, the least significant bit is bit B0, and
the most significant bit is bit B15. For example, if a value of
1.03
(which is 1026) is read as the value
of the condition register, the binary equivalent is 0000 0100 0000 0010. This value indicates that bit B1 and bit
B10 are set.
B15
B14
B13
B12
B11
B10
B9
B8
B7
B6
B5
B4
B3
B2
B1
B0
**
>
>
>
>
>
>
>
>
>
>
>
>
>
>
*
0
0
0
0
0
1
0
0
0
0
0
0
0
0
1
0
* Least significant bit
** Most significant bit
For information about .condition, .enable, .event, .ntr, and .ptr registers, refer to
Status register set contents
(on
page E-1) and
Enable and transition registers
(on page E-19). The individual bits of this register are defined in
the following table.
Bit
Value
Description
B0
status.operation.instrument.lan.CONNECTION
status.operation.instrument.lan.CON
Set bit indicates that the LAN cable
is connected and a link has been
detected.
Bit B0 decimal value: 1
B1
status.operation.instrument.lan.CONFIGURING
status.operation.instrument.lan.CONF
Set bit indicates the LAN is
performing its configuration
sequence.
Bit B1 decimal value: 2
B2-B9
Not used
Not available
B10
status.operation.instrument.lan.TRIGGER_OVERRUN
status.operation.instrument.lan.TRGOVR
Set bit indicates one or more
enabled bits for the operation status
LAN trigger overrun register is set.
Bit B10 decimal value: 1,024
B11-B15 Not used
Not applicable
As an example, to set bit B0 of the operation status LAN summary enable register, set
status.operation.instrument.lan.enable
= status.operation.instrument.lan.CON
.
In addition to the above constants,
operationRegister
can be set to the numeric equivalent of the bit to set.
To set more than one bit of the register, set
operationRegister
to the sum of their decimal weights. For
example, to set bits B1 and B10, set
operationRegister
to 1,026 (which is the sum of 2 + 1024).
Summary of Contents for System SourceMeter 2601B