Series 2600B System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2600BS-901-01 Rev. B / May 2013
7-231
smuX.source.settling
This attribute contains the source settling mode.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
SMU reset
Instrument reset
Recall setup
Not saved
0 (smuX.SETTLE_SMOOTH)
Usage
settleOption
= smu
X
.source.settling
smu
X
.source.settling =
settleOption
settleOption
Set to the source settling mode. Set to one of the following values:
0
or
smu
X
.SETTLE_SMOOTH
: Turns off additional settling operations (default)
1
or
smu
X
.SETTLE_FAST_RANGE
: Instructs the source-measure unit (SMU) to use
a faster procedure when changing ranges
2
or
smu
X
.SETTLE_FAST_POLARITY
: Instructs the SMU to change polarity without
going to zero
3
or
smu
X
.SETTLE_DIRECT_IRANGE
: Instructs the SMU to change the current
range directly
4
or
smu
X
.SETTLE_SMOOTH_100NA
: Enables the use of range rampers for the 100
nA range
128
or
smu
X
.SETTLE_FAST_ALL
: Enables all
smu
X
.SETTLE_FAST_*
operations
X
SMU channel (for example,
smua.source.settling
applies to SMU channel A)
Details
Using
smu
X
.SETTLE_FAST_RANGE
may cause the SMU to exceed the range change overshoot specification.
smu
X
.SETTLE_FAST_POLARITY
does not go to zero when changing polarity and may create inconsistencies at
the zero crossing.
smu
X
.SETTLE_DIRECT_IRANGE
switches the SMU directly to the target range instead of the default “range-by-
range” method. This option is mutually exclusive of any other
smu
X
.SETTLE_FAST_*
commands.
smu
X
.SETTLE_SMOOTH_100NA
is disabled by default in Model 2601B/2602B/2604B and Model
2611B/2612B/2614B. In Model 2634B/2635B/2636B, it is always enabled.
Example
smua.source.settling = smua.SETTLE_FAST_POLARITY
Selects fast polarity changing for
SMU channel A.
Also see
Range
(on page 2-82)
smuX.source.sink
This attribute turns sink mode on or off.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
SMU reset
Instrument reset
Recall setup
Saved setup
0 (smuX.DISABLE)