Section 7: TSP command reference
Series 2600B System SourceMeter® Instrument Reference Manual
7-154
2600BS-901-01 Rev. B / May 2013
When there are outstanding overlapped commands to acquire data,
n
refers to the index that the last entry in the
table will have after all the readings have completed.
If you pass a reading buffer instead of a reading buffer subtable, the default subtable for that reading buffer will
be used.
This command generates a single response message that contains all data. The response message is stored in
the output queue.
The
format.data
attribute controls the format of the response message.
The following bufferVar attributes can be used with the print buffer command.
bufferVar.dates, bufferVar.fillmode, bufferVar.formattedreadings, bufferVar.fractionalseonds, bufferVar.logstate,
bufferVar.seconds, buffer.units, bufferVar.relativetimestamps, bufferVar.seconds, bufferVar.sourcestatuses,
bufferVar, sourceunits, bufferVar.sourcevalues, bufferVar.statuses, bufferVar.times, bufferVar.timestamps,
bufferVar.units
If you are using the printbuffer() command to print multiple readings from multiple buffers, the readings must be
paired: buffer 1 readings, buffer 2 readings, next buffer 1 readings, next buffer 2 readings and so on. For
example, when you are storing voltage readings in one buffer and current readings in another buffer, keep them
paired in the response. Or, when you are just storing different readings in multiple buffer keep then paired in
response.