Series 2600B System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2600BS-901-01 Rev. B / May 2013
7-79
Also see
display.smuX.limit.func
(on page 7-77)
display.trigger.clear()
This function clears the front-panel trigger event detector.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
display.trigger.clear()
Details
The trigger event detector remembers if an event has been detected since the last
display.trigger.wait()
call. This function clears the trigger event detector and discards the previous history of TRIG key presses.
This attribute also clears the
display.trigger.overrun
attribute.
Also see
display.trigger.overrun
(on page 7-79)
display.trigger.wait()
(on page 7-80)
display.trigger.EVENT_ID
This constant is the event ID of the event generated when the front-panel TRIG key is pressed.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Constant
Yes
Usage
eventID
= display.trigger.EVENT_ID
eventID
The trigger event number
Details
Set the stimulus of any trigger event detector to the value of this constant to have it respond to front-panel trigger
key events.
Also see
None
display.trigger.overrun
This attribute contains the event detector overrun status.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (R)