Series 2600B System SourceMeter® Instrument Reference Manual
Section 2: General operation
2600BS-901-01 Rev. B / May 2013
2-97
LAN status LEDs
The figure below illustrates the two status light emitting diodes (LED) that are located at the top of the
LAN connection port of the instrument. The table below the figure provides explanations of the LED
states.
Figure 49: LAN Status
1 LED indicates the LAN port is connected to a 100 Mbps network
2 LED indicates the LAN port is connected to a 10 Mbps network
When the LED is:
The network:
Off
is not connected
On
is connected
Blinking
is sending or receiving data
Using the LAN with remote operations
The following table lists the Series 2600B remote interface's available LAN protocols:
LAN protocols
Port number
Protocol
23
Telnet
1024
VXI-11
5025
Raw socket
5030
Dead socket termination port
You can only use one remote interface at a time. Although multiple ethernet connections to the
instrument can be opened, only one can be used to control the instrument at a time.