Series 2600B System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2600BS-901-01 Rev. B / May 2013
7-131
Details
This read-only attribute is set to
true
when the LAN trigger is connected and ready to send trigger events
following a successful
lan.trigger[
N
].connect()
command; if the LAN trigger is not ready to send trigger
events, this value is
false
.
This attribute is also
false
when either
lan.trigger[
N
].protocol
or
lan.trigger[
N
].ipaddress
attributes are changed or the remote connection closes the connection.
Example
lan.trigger[1].protocol = lan.MULTICAST
print(lan.trigger[1].connected)
Outputs
true
if connected, or
false
if not
connected.
Example output:
false
Also see
lan.trigger[N].connect()
(on page 7-130)
lan.trigger[N].ipaddress
(on page 7-132)
lan.trigger[N].protocol
(on page 7-134)
lan.trigger[N].disconnect()
This function disconnects the LAN trigger.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
lan.trigger[
N
].disconnect()
N
The LAN event number (1 to 8)
Details
For TCP connections, this closes the TCP connection.
The LAN trigger automatically disconnects when either the
lan.trigger[
N
].protocol
or
lan.trigger[
N
].ipaddress
attributes for this event are changed.
Also see
lan.trigger[N].ipaddress
(on page 7-132)
lan.trigger[N].protocol
(on page 7-134)
lan.trigger[N].EVENT_ID
This constant is the event identifier used to route the LAN trigger to other subsystems (using stimulus properties).
Type
TSP-Link accessible
Affected by
Where saved
Default value
Constant
Yes
Usage
lan.trigger[
N
].EVENT_ID
N
The LAN event number (1 to 8)