Section 7: TSP command reference
Series 2600B System SourceMeter® Instrument Reference Manual
7-360
2600BS-901-01 Rev. B / May 2013
Example
print(tsplink.trigger[1].overrun)
If an event was ignored, displays
true
; if an event was not ignored,
displays
false
.
Also see
tsplink.trigger[N].assert()
(on page 7-356)
tsplink.trigger[N].clear()
(on page 7-356)
tsplink.trigger[N].mode
(on page 7-358)
tsplink.trigger[N].release()
tsplink.trigger[N].reset()
(on page 7-361)
tsplink.trigger[N].stimulus
tsplink.trigger[N].wait()
(on page 7-364)
tsplink.trigger[N].pulsewidth
This attribute sets the length of time that the trigger line is asserted for output triggers. This attribute is not
available on the Models 2604B/2614B/2634B.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Instrument reset
TSP-Link trigger N reset
Recall setup
Not saved
10e-6 (10
µ
s)
Usage
width
= tsplink.trigger[
N
].pulsewidth
tsplink.trigger[
N
].pulsewidth =
width
width
The pulse width (in seconds)
N
The trigger line (1 to 3)
Details
Setting the pulse width to 0 (seconds) asserts the trigger indefinitely.
Example
tsplink.trigger[3].pulsewidth = 20e-6
Sets pulse width for trigger line 3 to 20
μs.
Also see
tsplink.trigger[N].release()
tsplink.trigger[N].release()
This function releases a latched trigger on the given TSP-Link trigger line. This function is not available on the
Models 2604B/2614B/2634B.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
tsplink.trigger[
N
].release()
N
The trigger line (1 to 3)
Summary of Contents for System SourceMeter 2601B