Series 2600B System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2600BS-901-01 Rev. B / May 2013
7-333
Figure 138: SweepVLinMeasureI()
Example
SweepVLinMeasureI(smua, -1, 1, 1E-3, 1000)
This function performs a 1000-point linear
voltage sweep starting at -1 V and stopping at
+1 V. Current is measured at every step (point)
in the sweep after a 1 ms source settling period.
Also see
KISweep factory script
(on page 5-20)
SweepVListMeasureI()
This KISweep factory script function performs a voltage list sweep with current measured at every step (point).
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
No
Usage
SweepVListMeasureI(
smu, vlist, stime, points
)
smu
System SourceMeter
®
instrument channel (for example,
smua
refers to SMU channel
A)
vlist
Arbitrary list of voltage source values;
vlist = {value1, value2, ... valueN}
stime
Settling time in seconds; occurs after stepping the source and before performing a
measurement
points
Number of sweep points (must be
≥
2)
Details
Data for current measurements, voltage source values, and timestamps are stored in
smu
X
.nvbuffer1
.
If all parameters are omitted when this function is called, this function is executed with the parameters set to the
default values.
Performs a voltage list sweep with current measured at every step (point):
1. Sets the
smu
to output
vlist
volts value, allows the source to settle for
stime
seconds, and then
performs a current measurement.
2. Sets the
smu
to output the next
vlist
volts value, allows the source to settle for
stime
seconds, and
then performs a current measurement.
3. Repeats the above sequence until the current is measured for the last volts value. The last point in the
list to be measured is
points
.