Section 7: TSP command reference
Series 2600B System SourceMeter® Instrument Reference Manual
7-364
2600BS-901-01 Rev. B / May 2013
tsplink.trigger[N].wait()
This function waits for a trigger. This function is not available on the Models 2604B/2614B/2634B.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
triggered
= tsplink.trigger[
N
].wait(
timeout
)
triggered
Trigger detection indication; set to one of the following values:
true
: A trigger is detected during the timeout period
false
: A trigger is not detected during the timeout period
N
The trigger line (1 to 3)
timeout
The timeout value in seconds
Details
This function waits up to the timeout value for an input trigger. If one or more trigger events were detected since
the last time
tsplink.trigger[
N
].wait()
or
tsplink.trigger[
N
].clear()
was called, this function
returns immediately.
After waiting for a trigger with this function, the event detector is automatically reset and rearmed. This is true
regardless of the number of events detected.
Example
triggered = tsplink.trigger[3].wait(10)
print(triggered)
Waits up to 10 seconds for a trigger
on TSP-Link
®
line 3.
If
false
is returned, no trigger was
detected during the 10-second
timeout.
If
true
is returned, a trigger was
detected.
Also see
tsplink.trigger[N].clear()
(on page 7-356)
tsplink.writebit()
This function sets a TSP-Link synchronization line high or low. This function is not available on the Models
2604B/2614B/2634B.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
tsplink.writebit(
N
,
data
)
N
The trigger line (1 to 3)
data
The value to write to the bit:
•
Low: 0
•
High: 1
Summary of Contents for System SourceMeter 2601B