Appendix B: Calibration
Series 2600B System SourceMeter® Instrument Reference Manual
B-22
2600BS-901-01 Rev. B / May 2013
Step sequence
Adjustment steps must be performed in a specific sequence. See the following table that is specific
table to your model. Note that all steps are performed using 2-wire (local sensing) except as noted.
Adjustment of each range is performed as a four-point calibration:
•
+
ZERO
•
+
FULL SCALE
•
−
ZERO
•
−
FULL SCALE
Before performing the adjustment steps, refer to
Parameter values
(on page B-21) and
Sense modes
(on page B-21).
Model 2601B/2602B/2604B calibration steps
Function
1
Calibration steps
2
Calibration points
4
Sense mode
5
Voltage Source
and Measure
100 mV
100 mV
1 V
1 V
6 V
40 V
±1e-30, ±90 mV
±1e-30, ±90 mV
±1e-30, ±0.9 V
±1e-30, ±0.9 V
±1e-30, ±5.4 V
±1e-30, ±36 V
smu
X
.SENSE_LOCAL
smu
X
.SENSE_REMOTE
smu
X
.SENSE_LOCAL
smu
X
.SENSE_CALA
smu
X
.SENSE_LOCAL
smu
X
.SENSE_LOCAL
Current Source and Measure
100 nA
1 μA
10 μA
100 μA
1 mA
1 mA
10 mA
100 mA
1 A
3 A
10 A
3
±1e-30, ±90 nA
±1e-
30, ±0.9 μA
±1e-
30, ±9 μA
±1e-
30, ±90 μA
±1e-30, ±0.9 mA
±1e-30, ±0.9 mA
±1e-30, ±9 mA
±1e-30, ±90 mA
±1e-30, ±0.9 A
±1e-30, ±2.4 A
±1e-30, ±2.4 A
smu
X
.SENSE_LOCAL
smu
X
.SENSE_LOCAL
smu
X
.SENSE_LOCAL
smu
X
.SENSE_LOCAL
smu
X
.SENSE_LOCAL
smu
X
.SENSE_CALA
smu
X
.SENSE_LOCAL
smu
X
.SENSE_LOCAL
smu
X
.SENSE_LOCAL
smu
X
.SENSE_LOCAL
smu
X
.SENSE_LOCAL
1. Calibrate only the source for the CALA sense steps.
2. Steps must be performed in the order shown.
3. 10 A range for changing calibration of range only and is not available for normal use.
4. Do not use actual 0 values for zero calibration points. Send very small values such as ±1e-30. Calibration polarities must also be
set as shown in the procedures.
5. Output must be off before changing to the CALA sense mode.
Summary of Contents for System SourceMeter 2601B