Section 2: General operation
Series 2600B System SourceMeter® Instrument Reference Manual
2-86
2600BS-901-01 Rev. B / May 2013
Range programming
Range commands
The following tables summarize commands necessary to control measure and source ranges. See
the
TSP command reference
(on page 7-1) for more details about these commands.
Measure range commands*
Command**
Description
smu
X
.measure.autorangei = smu
X
.AUTORANGE_ON
Enable current measure autorange.
smu
X
.measure.autorangei = smu
X
.AUTORANGE_OFF
Disable current measure autorange.
smu
X
.measure.autorangev = smu
X
.AUTORANGE_ON
Enable voltage measure autorange.
smu
X
.measure.autorangev = smu
X
.AUTORANGE_OFF
Disable voltage measure autorange.
smu
X
.measure.lowrangei = lowrange
Set lowest I measure range for
autorange.
smu
X
.measure.lowrangev = lowrange
Set lowest V measure range for
autorange.
smu
X
.measure.rangei = rangeval
Select manual current measure range.
smu
X
.measure.rangev = rangeval
Select manual voltage measure range.
* See
Available ranges
**
smu
X
: For Models 2601B, 2611B, and 2635B, this value is
smua
(SMU Channel A); for Models 2602B, 2604B, 2612B,
2614B, 2634B, and 2636B, this value can be
smua
(for SMU Channel A) or
smub
(for SMU Channel B).
Source range and limit commands*
Command**
Description
smu
X
.source.autorangei = smu
X
.AUTORANGE_ON
Enable current source autorange.
smu
X
.source.autorangei = smu
X
.AUTORANGE_OFF
Disable current source autorange.
smu
X
.source.autorangev = smu
X
.AUTORANGE_ON
Enable voltage source autorange.
smu
X
.source.autorangev = smu
X
.AUTORANGE_OFF
Disable voltage source autorange.
smu
X
.source.limiti = level
Set voltage source current limit
(compliance).
smu
X
.source.limitv = level
Set current source voltage limit
(compliance).
smu
X
.source.limitp = level
Set source power limit (compliance).
smu
X
.source.lowrangei = lowrange
Set lowest I source range for autorange.
smu
X
.source.lowrangev = lowrange
Set lowest V source range for autorange.
smu
X
.source.rangei = rangeval
Select manual current source range.
smu
X
.source.rangev = rangeval
Select manual voltage source range.
* See
Available ranges
**
smu
X
: For Models 2601B, 2611B, and 2635B, this value is
smua
(SMU Channel A); for Models 2602B, 2604B, 2612B,
2614B, 2634B, and 2636B, this value can be
smua
(for SMU Channel A) or
smub
(for SMU Channel B).
Summary of Contents for System SourceMeter 2601B
Page 841: ......