Series 2600B System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2600BS-901-01 Rev. B / May 2013
7-217
smuX.sense
This attribute contains the state of the sense mode.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
SMU reset
Instrument reset
Recall setup
Saved setup
0 (smuX.SENSE_LOCAL)
Usage
senseMode
= smu
X
.sense
smu
X
.sense =
senseMode
senseMode
The sense mode; set to one of the following:
0
or
smu
X
.SENSE_LOCAL
: Selects local sense (2-wire)
1
or
smu
X
.SENSE_REMOTE
: Selects remote sense (4-wire)
3
or
smu
X
.SENSE_CALA
: Selects calibration sense mode
X
Source-measure unit (SMU) channel (for example,
smua.sense
applies to SMU
channel A)
Details
Source-measure operations are performed using either 2-wire local sense connections or 4-wire remote sense
connections. Writing to this attribute selects the sense mode.
The
smu
X
.SENSE_CALA
mode is only used for calibration and may only be selected when calibration is enabled.
The sense mode can be changed between local and remote while the output is on.
The calibration sense mode cannot be selected while the output is on.
Resetting the instrument selects the local sense mode.
Example
smua.sense = smua.SENSE_REMOTE
Selects remote sensing for SMU
channel A.
Also see
2-wire local sensing connections
(on page 2-53)
4-wire remote sensing connections
(on page 2-54)
Sense mode selection
(on page 2-76)
smuX.source.autorangeY
This attribute contains the state of the source autorange control (on/off).
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
SMU reset
Instrument reset
Recall setup
Saved setup
1 (smuX.AUTORANGE_ON)
Summary of Contents for System SourceMeter 2601B
Page 841: ......