Sense mode selection
The sense mode can be set to use
2-wire local sensing connections
(on page 2-53) or
4-wire remote
sensing connections
(on page 2-54).
The default sense setting is 2-wire local.
Front panel sense mode selection
To check or change the sense mode from the front panel:
1. Press the
CONFIG
key.
2. Press the
SRC
or
MEAS
key. You can access and set the Series 2600B sense mode from either
the V-SOURCE or the V-MEAS menu items.
3. If you pressed the SRC key: select
V-SOURCE
>
SENSE-MODE
, and then press the
ENTER
key
or the navigation wheel
.
If you pressed the MEAS key: select
V-MEAS
>
SENSE-MODE
, and then press the
ENTER
key
or the navigation wheel
.
4. Select
2-WIRE
or
4-WIRE
as needed, and then press the
ENTER
key or the navigation wheel
.
Selecting the sense from the remote interface
To select the remote sense from the remote interface:
Set the
smu
X
.sense
attribute to control the sense state by remote. The programming example below
illustrates how to configure the Series 2600B for 4-wire remote sensing:
smua.sense = smua.SENSE_REMOTE
The following table summarizes the commands to select the sense mode. See
Remote source-
measure commands
(on page 2-34) and
TSP command reference
(on page 7-1) for details on using
these commands.
Commands to select sense mode
Command*
Description
smu
X
.source.output = smu
X
.OUTPUT_OFF
Turns off the source-measure unit
(SMU) output.
smu
X
.sense = smu
X
.SENSE_LOCAL
Select local (2-wire) sense.
smu
X
.sense = smu
X
.SENSE_LOCAL
Select remote (4-wire) sense.
*
smu
X
: For Models 2601B, 2611B, and 2635B, this value is
smua
(SMU Channel A); for Models 2602B,
2604B, 2612B, 2614B, 2634B, and 2636B, this value can be
smua
(for SMU Channel A) or
smub
(for SMU
Channel B).