Series 2600B System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2600BS-901-01 Rev. B / May 2013
7-179
Example
settimezone("8", "1", "3.3.0/02", "11.2.0/02")
settimezone(offset)
Sets
offset
to equal +8 hours, +1
hour for DST, starts on Mar 14 at 2:00
a.m, ends on Nov 7 at 2:00 a.m.
Sets local time zone to
offset
.
Also see
gettimezone()
(on page 7-98)
os.time()
(on page 7-151)
settime()
(on page 7-177)
setup.poweron
This attribute specifies which saved setup to recall when the instrument is turned on.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Not applicable
Nonvolatile memory
0
Usage
id
= setup.poweron
setup.poweron =
id
id
An integer that specifies the setup to recall when the instrument power is
turned on (0 to 5)
Details
When
id
=
0
, the instrument uses the factory default setup when it is turned on. When
id
is set to 1 to 5, it uses
the setup saved with
setup.save()
.
Only setups stored in nonvolatile memory are available (you cannot recall a script from a USB flash drive with
this command).
Example
setup.poweron = 0
Set the instrument to use the factory default setup
when power is turned on.
Also see
setup.save()
(on page 7-181)
Start-up (power-on) configuration
(on page 2-48)