Series 2600B System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2600BS-901-01 Rev. B / May 2013
7-357
Example
tsplink.trigger[2].clear()
Clears trigger event on synchronization line 2.
Also see
tsplink.trigger[N].mode
(on page 7-358)
tsplink.trigger[N].overrun
(on page 7-359)
tsplink.trigger[N].release()
tsplink.trigger[N].stimulus
tsplink.trigger[N].wait()
(on page 7-364)
tsplink.trigger[N].EVENT_ID
This constant identifies the number that is used for the trigger events. This constant is not available on the Models
2604B/2614B/2634B.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Constant
Yes
Usage
eventID
= tsplink.trigger[
N
].EVENT_ID
eventID
The trigger event number
N
The trigger line (1 to 3)
Details
This number is used by the TSP-Link trigger line when it detects an input trigger.
Set the stimulus of any trigger object to the value of this constant to have the trigger object respond to trigger
events from this line.
Example
trigger.timer[1].stimulus = tsplink.trigger[2].EVENT_ID
Sets the trigger stimulus
of trigger timer 1 to the
TSP-Link trigger 2 event.
Also see
None