Section 7: TSP command reference
Series 2600B System SourceMeter® Instrument Reference Manual
7-356
2600BS-901-01 Rev. B / May 2013
tsplink.trigger[N].assert()
This function simulates the occurrence of the trigger and generates the corresponding event ID. This function is
not available on the Models 2604B/2614B/2634B.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
tsplink.trigger[
N
].assert()
N
The trigger line (1 to 3)
Details
The set pulse width determines how long the trigger is asserted.
Example
tsplink.trigger[2].assert()
Asserts trigger on trigger line 2.
Also see
tsplink.trigger[N].clear()
(on page 7-356)
tsplink.trigger[N].mode
(on page 7-358)
tsplink.trigger[N].overrun
(on page 7-359)
tsplink.trigger[N].pulsewidth
(on page 7-360)
tsplink.trigger[N].release()
tsplink.trigger[N].stimulus
tsplink.trigger[N].wait()
(on page 7-364)
tsplink.trigger[N].clear()
This function clears the event detector for a trigger. This function is not available on the Models
2604B/2614B/2634B.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
tsplink.trigger[
N
].clear()
N
The trigger line (1 to 3)
Details
The trigger event detector enters the detected state when an event is detected.
tsplink.trigger[
N
].clear()
clears a trigger event detector, discards the history of the trigger line, and
clears the
tsplink.trigger[
N
].overrun
attribute.
Summary of Contents for System SourceMeter 2601B