TI00081D/06/EN/13.11
71141466

Technical Information

Proline Promag 10D

Electromagnetic flow measuring system
Flow measurement of liquids in water or wastewater applications

Application

Electromagnetic flowmeter for bidirectional 
measurement of liquids with:
• A minimum conductivity of 



50 

μ

S/cm

– Drinking water
– Wastewater

• Flow measurement up to 4700 dm³/min 

(1250 gal/min)

• Fluid temperature up to +60 °C (+140 °F)

• Process pressures up to 16 bar (232 psi)

• Polyamide lining material

• Drinking water approvals:

– KTW/W270
– WRAS BS 6920
– ACS
– NSF 61

Your benefits

The measuring devices offer you cost-effective flow 
measurement with a high degree of accuracy for a wide 
range of process conditions.

The uniform Proline transmitter concept comprises:
• High degree of reliability and measuring stability
• Uniform operating concept

The tried-and-tested Promag sensors offer:
• No pressure loss
• Not sensitive to vibrations
• Simple installation and commissioning

Summary of Contents for Proline Promag 10D

Page 1: ... min Fluid temperature up to 60 C 140 F Process pressures up to 16 bar 232 psi Polyamide lining material Drinking water approvals KTW W270 WRAS BS 6920 ACS NSF 61 Your benefits The measuring devices offer you cost effective flow measurement with a high degree of accuracy for a wide range of process conditions The uniform Proline transmitter concept comprises High degree of reliability and measurin...

Page 2: ...g kit 14 Length of connecting cable 14 Environment 15 Ambient temperature range 15 Storage temperature 15 Degree of protection 15 Shock and vibration resistance 15 Electromagnetic compatibility EMC 15 Process 15 Medium temperature range 15 Medium pressure range 15 Conductivity 15 Pressure temperature ratings 15 Pressure tightness 16 Limiting flow 16 Pressure loss 16 Mechanical construction 17 Desi...

Page 3: ... of the pipe cross sectional area The DC magnetic field is created through a switched direct current of alternating polarity a0003191 Ue B L v Q A v Ue Induced voltage B Magnetic induction magnetic field L Electrode spacing v Flow velocity Q Volume flow A Pipe cross section I Current strength Measuring system The measuring system consists of a transmitter and a sensor Two versions are available Co...

Page 4: ...ctor Can be configured as Pulse output Pulse value and pulse polarity can be selected max pulse width adjustable 5 to 2000 ms pulse frequency max 100 Hz Status output For example can be configured for error messages flow recognition limit value Signal on alarm Current output Failsafe mode can be selected e g in accordance with NAMUR Recommendation NE 43 Pulse output Failsafe mode can be selected S...

Page 5: ...PROM saves measuring system data Electrical connection A0003192 Connecting the transmitter cable cross section max 2 5 mm2 14 AWG a Electronics compartment cover b Power supply cable c Ground terminal for protective ground d Terminal connector for power supply cable e Signal cable f Ground terminal for signal cable g Terminal connector for signal cable h Service connector i Ground terminal for pot...

Page 6: ...wing should also be taken into consideration for potential equalization Internal grounding concepts in the company Operating conditions such as the material grounding of the pipes cathodic protection etc see table Standard b S1 E1 E2 S2 GND E S 42 41 6 5 7 8 4 37 36 d c a d c E1 E2 GND E 5 7 4 37 42 41 n c n c Operating conditions Potential equalization When using the measuring device in a Metal g...

Page 7: ... This connection method also applies in situations where Customary potential equalization cannot be ensured Excessively high equalizing currents can be expected Potential equalization takes place via the ground terminal of the transmitter and the two pipe flanges Here the ground cable copper wire 6 mm2 10 AWG is mounted directly on the conductive flange coating with flange screws A0010703 Via the ...

Page 8: ...trical interference The measuring device complies with the general safety requirements in accordance with EN 61010 1 the EMC requirements of IEC EN 61326 and NAMUR Recommendation NE 21 Caution Grounding is by means of the ground terminals provided for the purpose inside the connection housing Ensure that the stripped and twisted lengths of cable shield to the ground terminal are as short as possib...

Page 9: ...utlet run 5 DN Sensor and transmitter grounded The sensor is centered in the pipe Maximum measured error Current output also typically 5 μA Pulse output 0 5 o r 2 mm s 0 5 o r 0 08 in s o r of reading Fluctuations in the supply voltage do not have any effect within the specified range A0003200 Max measured error in of reading Repeatability Max 0 2 o r 2 mm s 0 2 o r 0 08 in s o r of reading 2 5 2 ...

Page 10: ...k of air accumulating Directly upstream from a free pipe outlet in a vertical pipeline Risk of pipe not filling correctly A0010705 Mounting location Orientation h 2 x DN Vertical Horizontal Vertical orientation is generally preferred Vertical orientation helps avoid gas and air accumulations and deposits in the measuring tube A0010709 Vertical orientation The measuring electrode axis should be hor...

Page 11: ...ion resistance of the device ä 15 section Shock and vibration resistance A0010706 Installing the device with a pump Partially filled pipes Partially filled pipes with gradients necessitate a drain type configuration A0010707 Installation with partially filled pipes Down pipes Install a siphon or a vent valve downstream of the sensor in down pipes h 5 m 16 4 ft This precaution is to avoid low press...

Page 12: ...hock and vibration resistance ä 15 section Shock and vibration resistance A0010711 Measures to prevent vibration of the measuring device L 10 m 32 8 ft Inlet and outlet runs If possible install the sensor well clear of assemblies such as valves T pieces elbows etc The following inlet and outlet runs must be observed in order to meet accuracy specifications Inlet run 5 DN Outlet run 2 DN A0010712 I...

Page 13: ...can be used to calculate the pressure loss caused by reducers and expanders Note The nomogram only applies to liquids of viscosity similar to water The pressure loss is calculated as follows 1 Calculate the diameter ratio d D 2 Read off the pressure loss as a function of flow velocity downstream from the reduction and the d D ratio from the nomogram A0010713 Pressure loss due to adapters 100 10 0 ...

Page 14: ...e version Fix cable run or lay in armored conduit Note Cable movements can falsify the measuring signal especially in the case of low fluid conductivities Route the cable well clear of electrical machines and switching elements If necessary establish potential equalization between the sensor and transmitter The permitted cable length Lmax is determined by the fluid conductivity A minimum conductiv...

Page 15: ...re moisture does not collect in the measuring device This will help prevent fungus and bacteria infestation which can damage the liner Degree of protection IP 67 NEMA 4X for transmitter and sensor Shock and vibration resistance Acceleration up to 2 g following IEC 600 68 2 6 Electromagnetic compatibility EMC As per IEC EN 61326 and NAMUR Recommendation NE 21 Emission to limit value for industry EN...

Page 16: ...orating adapters according to DIN EN 545 ä 13 section Adapters Flow characteristic values SI units Diameter Recommended flow Factory settings mm inch min max full scale value v 0 3 bzw 10 m s dm3 min Full scale value current output v 2 5 m s dm3 min Pulse value 2 pulses s dm3 Low flow cut off v 0 04 m s dm3 min 25 1 9 to 300 75 0 50 1 40 1 25 to 700 200 1 50 3 50 2 35 to 1100 300 2 50 5 65 60 to 2...

Page 17: ...06 65 93 272 70 342 139 60 125 80 117 276 75 351 151 76 135 100 148 290 89 379 179 97 160 All dimensions in mm DN A B C D E F G H J d 1 d 2 e 1 e 2 ANSI max seals 1 7 00 0 79 to 1 08 6 02 to 6 61 4 45 2 17 5 90 9 45 1 69 11 1 3 39 0 94 2 68 1 2 72 9 88 2 05 11 9 4 11 1 50 3 43 2 3 27 10 3 2 44 12 8 4 88 1 97 4 17 3 4 61 10 9 2 95 13 8 5 94 2 99 5 43 4 5 83 11 4 3 50 14 9 7 05 3 82 6 30 All dimensi...

Page 18: ... K F E D A H L N O P J G B C M ANSCHLUSSKLEMMEN FIELD TERMINALS A B C D E F G Ø H 178 113 135 20 to 27 5 153 to 168 113 100 8 6 ISO M8 J K L M N O P 123 150 100 25 133 177 5 327 5 All dimensions in mm A B C D E F G Ø H 7 00 14 5 5 31 0 79 to 1 08 6 02 to 6 61 4 45 3 94 0 34 ISO M8 J K L M N O P 4 84 5 90 3 94 0 98 5 24 6 99 12 9 All dimensions in inch ...

Page 19: ...omag 10D Endress Hauser 19 A0010719 Transmitter mounting remote version A Direct wall mounting B Pipe mounting 248 2 9 76 0 08 mm inch 238 9 37 A B ANSCHLUSSKLEMMEN FIELD TERMINALS ANSCHLUSSKLEMMEN FIELD TERMINALS ...

Page 20: ...294 139 60 125 80 117 228 75 303 151 76 135 100 148 242 89 331 179 97 160 All dimensions in mm DN A B C D E F G H d 1 d 2 e 1 e 2 ANSI max seals 1 5 08 6 42 5 63 2 17 4 02 7 56 1 69 9 25 3 39 0 94 2 68 1 2 72 7 99 2 05 10 0 4 11 1 50 3 43 2 3 27 8 43 2 44 10 9 4 88 1 97 4 17 3 4 61 8 98 2 95 11 9 5 94 2 99 5 43 4 5 83 9 53 3 50 13 0 7 05 3 82 6 30 Alle dimensions in inch E C B A DN 25 40 1 1 DN 50...

Page 21: ...er inch mm inch mm inch mm inch mm 1 25 EN DIN PN16 4 M12 5 71 145 2 13 54 0 94 24 1 40 4 M16 6 69 170 2 68 68 1 50 38 2 50 4 M16 7 28 185 3 23 82 1 97 50 65 4 M16 7 87 200 3 62 92 2 36 60 65 8 M16 7 87 200 2 36 60 3 80 8 M16 8 86 225 4 57 116 2 99 76 4 100 8 M16 10 24 260 5 79 147 3 82 97 EN DIN flange 4 hole with centering sleeves EN DIN flange 8 hole without centering sleeves A centering sleeve...

Page 22: ...N 1092 1 DIN 2501 ANSI B16 5 JIS B2220 Operability Local operation Display elements Liquid crystal display unilluminated two line 16 characters per line Display operating mode preconfigured volume flow and totalizer status 1 totalizer Operating elements Local operation via three keys Remote operation Operation via HART protocol and FieldCare Diameter Pressure rating Mounting bolts Centeringsleeves...

Page 23: ...by housing IP code EN 61010 1 Safety requirements for electrical equipment for measurement control and laboratory use IEC EN 61326 Emission in accordance with requirements for Class A Electromagnetic compatibility EMC requirements ANSI ISA S82 01 Safety Standard for Electrical and Electronic Test Measuring Controlling and related Equipment General Requirements Pollution degree 2 Installation Categ...

Page 24: ... power supply Limit relay Analog output RIA45 Process display RIA251 Digital display device for looping into the 4 to 20 mA current loop RIA251 Field display unit RIA16 Digital field display device for looping into the 4 to 20 mA current loop RIA16 Memograph M graphic display recorder The Memograph M graphic display recorder provides information on all the relevant process variables Measured value...

Page 25: ...ion via modem Ethernet or GSM Visualization via Internet Intranet in Web browser and or WAP cellular phone Limit value monitoring with alarm by e mail or SMS Synchronized time stamping of all measured values Remote diagnosis and remote configuration of connected HART devices FXA520 Accessory Description Order code Applicator Software for selecting and planning flowmeters The Applicator software ca...