Calibration
19-13
Table 19-5
20pA-2µA range amps calibration summary
Model 6514 range
Calibrator voltages
Standard resistor
1
Calibration currents
2
Settling time
3
20pA
100G
Ω
0pA
> 2 min.
+2.000000V
100G
Ω
+20pA
> 2 min.
-2.000000V
100G
Ω
-20pA
> 2 min.
200pA
10G
Ω
0pA
> 2 min.
+2.000000V
10G
Ω
+200pA
> 2 min.
-2.000000V
10G
Ω
-200pA
> 2 min.
2nA
1G
Ω
0nA
> 15 sec.
+2.000000V
1G
Ω
+2nA
> 15 sec.
-2.000000V
1G
Ω
-2nA
> 15 sec.
20nA
1G
Ω
0nA
> 15 sec.
+20.00000V
1G
Ω
+20nA
> 15 sec.
-20.00000V
1G
Ω
-20nA
> 15 sec.
200nA
100M
Ω
0nA
> 15 sec.
+20.00000V
100M
Ω
+200nA
> 15 sec.
-20.00000V
100M
Ω
-200nA
> 15 sec.
2
µ
A
100M
Ω
0
µ
A
> 15 sec.
+200.0000V
100M
Ω
+2µ
A
> 15 sec.
-200.0000V
100M
Ω
-2
µ
A
> 15 sec.
1
Nominal resistance values.
2
Nominal currents. Calculate actual currents from calibrator voltage and actual standard resistor value: I = V/R. Calibrate zero, posi-
tive full-scale, and negative full-scale for each range. Triax cap used for zero cal points.
3
Allow calibration signal to settle for indicated time before calibrating each point.
Summary of Contents for 6514
Page 203: ...14 CommonCommands...
Page 207: ...15 SCPISignalOriented MeasurementCommands...
Page 235: ...18 PerformanceVerification...
Page 254: ...19 Calibration...
Page 274: ...20 RoutineMaintenance...
Page 278: ...A Specifications...
Page 282: ...B StatusandErrorMessages...
Page 288: ...C GeneralMeasurement Considerations...
Page 294: ...D DDCEmulationCommands...
Page 299: ...E ExamplePrograms...
Page 306: ...F IEEE 488BusOverview...
Page 320: ...G IEEE 488andSCPI ConformanceInformation...
Page 324: ...H CalibrationOptions...