Calibration
19-5
Calibration errors
Model 6514 checks for errors after each calibration step, minimizing the possibility that
improper calibration may occur due to operator error. If an error is detected during calibration,
the instrument will display an appropriate error message. The unit will then prompt you to repeat
the calibration step that caused the error.
Calibration menu
You can access the calibration menu by pressing SHIFT then CAL. The various selections are
summarized in Table 19-2. Use the up or down RANGE key to scroll through these selections.
Aborting calibration
You can abort the calibration procedure at any time by pressing the EXIT key.
Table 19-2
Calibration menu
Menu item*
Description
RUN
Calibrates present function and range.
COUNT
Displays calibration count.
IOFFSET
Performs input bias current calibration.
VOFFSET
Performs offset voltage calibration.
DATES
Displays calibration and due dates.
UNLOCK
Unlocks calibration using code.
LOCK
Locks cal, exits to the main menu.
SAVE
Saves calibration constants.
* Press SHIFT then CAL to access. Use up or down RANGE to scroll through selections.
Summary of Contents for 6514
Page 203: ...14 CommonCommands...
Page 207: ...15 SCPISignalOriented MeasurementCommands...
Page 235: ...18 PerformanceVerification...
Page 254: ...19 Calibration...
Page 274: ...20 RoutineMaintenance...
Page 278: ...A Specifications...
Page 282: ...B StatusandErrorMessages...
Page 288: ...C GeneralMeasurement Considerations...
Page 294: ...D DDCEmulationCommands...
Page 299: ...E ExamplePrograms...
Page 306: ...F IEEE 488BusOverview...
Page 320: ...G IEEE 488andSCPI ConformanceInformation...
Page 324: ...H CalibrationOptions...