Performance Verification
18-19
7.
Verify charge measurement accuracy for each of the values listed in Table 18-7. For each
test point:
•
Select the correct Model 6514 measurement range.
•
Make connections to the correct standard capacitor.
•
Calculate the required voltage from the desired charge and actual standard capaci-
tance value: V = Q/C.
•
Disable zero check.
•
Set the calibrator voltage to the calculated value.
•
Verify that Model 6514 charge reading is within the required limits.
•
Enable zero check to discharge the capacitor.
Table 18-7
Coulombs measurement accuracy reading limits
Model
6514 range
Nominal
calibrator
voltage
Standard
capacitor
1
Applied charge
Actual
voltage
2
Model 6514 coulombs reading
limits (1 Year, 18˚C-28˚C)
20nC
20V
1nF
20.0000nC
_______V
19.9915 to 20.0805nC
200nC
200V
1nF
200.000nC
_______V
199.195 to 200.805nC
2µC
20V
100nF
2.00000µC
_______V
1.97995 to 202.02005µC
20
µ
C
200V
100nF
20.0000
µ
C
_______V
19.7995 to 20.2005
µ
C
1
Nominal values.
2
Calculate actual voltage from applied charge and actual capacitance value: V = Q/C.
Summary of Contents for 6514
Page 203: ...14 CommonCommands...
Page 207: ...15 SCPISignalOriented MeasurementCommands...
Page 235: ...18 PerformanceVerification...
Page 254: ...19 Calibration...
Page 274: ...20 RoutineMaintenance...
Page 278: ...A Specifications...
Page 282: ...B StatusandErrorMessages...
Page 288: ...C GeneralMeasurement Considerations...
Page 294: ...D DDCEmulationCommands...
Page 299: ...E ExamplePrograms...
Page 306: ...F IEEE 488BusOverview...
Page 320: ...G IEEE 488andSCPI ConformanceInformation...
Page 324: ...H CalibrationOptions...