6-10
Range, Units, Digits, Rate, and Filters
Operation consideration
•
The digital filter operation will reset (start over) whenever the zero check operation is
performed or the function is changed.
Digital filter configuration and control
The AVG key is a toggle-action key. It will either disable the digital filter (display “AVER-
AGE OFF”), or access the configuration menu to enable the digital filter.
NOTE
For the following procedure, use the
and
keys and the RANGE
and
keys to
set values. The
and
keys provide cursor control, while the
and
keys increment
and decrement the value.
1.
Press the AVG key to display the present number of reading conversions to average (filter
count).
2.
Key in desired filter count value (1 to 100) and press ENTER. The present filter type
(repeating or moving) is displayed.
3.
Use the
or
key to display the desired filter type (REPEAT or MOVNG AV) and press
ENTER.
SCPI programming — filters
Table 6-4
SCPI commands — filters
Commands
Description
Default
For median filter:
[:SENSe[1]]
SENSe Subsystem:
:MEDian
Median Filter:
:RANK <n>
Specify filter rank; 1 to 5.
1
[:STATe] <b>
Enable or disable median filter.
OFF
For digital filter:
[:SENSe[1]]
SENSe Subsystem:
:AVERage
Digital Filter:
:TCONtrol <name>
Select filter control; MOVing or REPeat.
REP
:COUNt <n>
Specify filter count; 1 to 100.
10
[:STATe] <b>
Enable or disable digital filter.
OFF
Summary of Contents for 6514
Page 203: ...14 CommonCommands...
Page 207: ...15 SCPISignalOriented MeasurementCommands...
Page 235: ...18 PerformanceVerification...
Page 254: ...19 Calibration...
Page 274: ...20 RoutineMaintenance...
Page 278: ...A Specifications...
Page 282: ...B StatusandErrorMessages...
Page 288: ...C GeneralMeasurement Considerations...
Page 294: ...D DDCEmulationCommands...
Page 299: ...E ExamplePrograms...
Page 306: ...F IEEE 488BusOverview...
Page 320: ...G IEEE 488andSCPI ConformanceInformation...
Page 324: ...H CalibrationOptions...