Status Structure
13-17
Event enable registers
As Figure 13-1 shows, each status register set has an enable register. Each event register bit
is logically ANDed (&) to a corresponding enable bit of an enable register. Therefore, when an
event bit is set and the corresponding enable bit is set (as programmed by the user), the output
(summary) of the register will set to 1, which in turn sets the summary bit of the status byte
register.
The commands to program and read the event enable registers are listed in Table 13-6. For
details on programming and reading registers, see “Programming enable registers” and “Read-
ing registers”.
NOTE
The bits of any enable register can be reset to 0 by sending the 0 parameter value
with the appropriate enable command (i.e. STATus:OPERation:ENABle 0).
Table 13-6
Common and SCPI commands — event enable registers
Command
Description
*ESE <NDN> or <NRf>
Program standard event enable register (see “Parameters”).
*ESE?
Read standard event enable register.
STATus
STATus subsystem:
:OPERation
Operation event enable register:
:ENABle <NDN> or <NRf>
Program enable register (see “Parameters”).
:ENABle?
Read enable register.
:MEASurement
Measurement event enable register:
:ENABle <NDN> or <NRf>
Program enable register (see “Parameters”).
:ENABle?
Read enable register.
:QUEStionable
Questionable event enable register:
:ENABle <NDN> or <NRf>
Program enable register (see “Parameters”).
:ENABle?
Read measurement event enable register:
Parameters:
<NDN> = #Bxx…x
Binary format (each x = 1 or 0)
= #Hx
Hexadecimal format (x = 0 to FFFF)
= #Qx
Octal format (x = 0 to 177777)
<NRf>
= 0 to 65535
Decimal format
Note: Power-up and STATus:PRESet resets all bits of all enable registers to 0. *CLS has no effect.
Summary of Contents for 6514
Page 203: ...14 CommonCommands...
Page 207: ...15 SCPISignalOriented MeasurementCommands...
Page 235: ...18 PerformanceVerification...
Page 254: ...19 Calibration...
Page 274: ...20 RoutineMaintenance...
Page 278: ...A Specifications...
Page 282: ...B StatusandErrorMessages...
Page 288: ...C GeneralMeasurement Considerations...
Page 294: ...D DDCEmulationCommands...
Page 299: ...E ExamplePrograms...
Page 306: ...F IEEE 488BusOverview...
Page 320: ...G IEEE 488andSCPI ConformanceInformation...
Page 324: ...H CalibrationOptions...