19-20
Calibration
Displaying the calibration count
To display the calibration count at any time:
1.
From normal display, press SHIFT then CAL. The unit will display the following:
CAL: RUN
2.
Use either RANGE key to select CAL:COUNT from the calibration menu, then press
ENTER. For example:
COUNT: 1
3.
Press EXIT to return to normal display.