5-4
Coulombs Measurements
Step 6
Connect the charge to be measured to the electrometer
Basic connections for amps measurements are shown in Figure 5-1.
NOTE
See “Connection Basics” in Section 2 for fundamental information on making con-
nections to the electrometer input.
Step 7
Take the charge reading from the display
If using auto discharge, use the REL key to zero the display when the integrator resets.
Remember that Rel was enabled in Step 5. Therefore, you will have to press REL twice. The first
press disables Rel, and the second press re-enables it to zero the display. See “Auto Discharge
Hop” in “Coulombs Measurement Considerations” (in this section).
Black (LO)
Red (HI)
237-ALG-2
Cable
Metal Noise Shield
(Optional)
6514 Rear Panel
Input LO connected
to shield
RS232
DIGITAL I/O
PREAMP OUT
250V PK
2V ANALOG
OUTPUT
COMMON
CHASSIS
120
FUSE
LINE
630mA
LINE RATING
50, 60Hz
60 VA MAX
T
(SB)
100 VAC
120 VAC
315mAT
(SB)
220 VAC
240 VAC
INPUT 250V PK
IEEE-488
(CHANGE IEEE ADDRESS
WITH FRONT PANEL MENU)
TRIGGER LINK
!
!
MADE IN
U.S.A.
V,
GUARD
(PROGRAMMABLE)
OFF
ON
GUARD
(FOLLOWS
INPUT)
(INTERNAL)
INPUT
PREAMP
10K
PREAMP
OUT
2V ANALOG
OUTPUT
COM
Figure 5-1
Typical connections
for coulombs
Summary of Contents for 6514
Page 203: ...14 CommonCommands...
Page 207: ...15 SCPISignalOriented MeasurementCommands...
Page 235: ...18 PerformanceVerification...
Page 254: ...19 Calibration...
Page 274: ...20 RoutineMaintenance...
Page 278: ...A Specifications...
Page 282: ...B StatusandErrorMessages...
Page 288: ...C GeneralMeasurement Considerations...
Page 294: ...D DDCEmulationCommands...
Page 299: ...E ExamplePrograms...
Page 306: ...F IEEE 488BusOverview...
Page 320: ...G IEEE 488andSCPI ConformanceInformation...
Page 324: ...H CalibrationOptions...