2-14
Measurement Concepts
For volts, amps and ohms, leave zero check enabled when connecting or disconnecting input
signals. For coulombs, disable zero check before connecting the input signal. If zero check is
left enabled when you connect the input signal, the charge will dissipate through the 10M
Ω
resistor (see Figure 2-10).
Zero correct
Model 6514 has a zero correct feature to algebraically subtract the voltage offset term from
the measurement. Perform the following steps to zero correct the volts or amps function:
NOTE
The ZCOR key toggles zero correct on and off. If zero correct is enabled (“ZZ” or
“CZ” message displayed), press ZCOR to disable it.
1.
Select the volts (V) or amps (I) function.
2.
Enable zero check (“ZC” message displayed).
3.
Select the range that will be used for the measurement, or select the lowest range.
4.
Press ZCOR to enable zero correct (“ZZ” message displayed).
5.
Press ZCHK to disable zero check.
6.
Readings can now be taken from the display. The “CZ” message indicates that the dis-
played reading is zero corrected.
C
IN
C
IN
= 10pF
Input
C
IN
C
IN
= 10pF
Input
C
IN
C
IN
= 10pF
Input
C
IN
Input
C
IN
= 10pF
C
F
= 1000pF (20nC, 200nC)
0.1
µ
F (2
µ
C, 20
µ
C)
10M
Ω
Volts and Ohms
Z
F
= 1k
Ω
(mA)
1M
Ω
|| 1000pF (
µ
A)
1G
Ω
|| 10pF (nA)
1T
Ω
|| 1pF (pA)
10M
Ω
Amps
Z
F
10M
Ω
Coulombs
10M
Ω
Ohms
Z
F
I
IN
C
F
Z
F
= 1k
Ω
(k
Ω
)
1M
Ω
|| 1000pF (M
Ω
)
1G
Ω
|| 10pF (G
Ω
)
Figure 2-10
Equivalent input
impedance with
zero check enabled
Summary of Contents for 6514
Page 203: ...14 CommonCommands...
Page 207: ...15 SCPISignalOriented MeasurementCommands...
Page 235: ...18 PerformanceVerification...
Page 254: ...19 Calibration...
Page 274: ...20 RoutineMaintenance...
Page 278: ...A Specifications...
Page 282: ...B StatusandErrorMessages...
Page 288: ...C GeneralMeasurement Considerations...
Page 294: ...D DDCEmulationCommands...
Page 299: ...E ExamplePrograms...
Page 306: ...F IEEE 488BusOverview...
Page 320: ...G IEEE 488andSCPI ConformanceInformation...
Page 324: ...H CalibrationOptions...