![background image](http://html2.mh-extra.com/html/keithley/6514/6514_instruction-manual_4064199227.webp)
SCPI Reference Tables
17-7
Command
Description
Default
parameter
Ref
SCPI
:ULIMit <NRf>
Select autorange upper limit; -0.021 to
0.021 (amps).
2.1e-2
:ULIMit?
Query upper limit for autorange.
:LLIMit <NRf>
Select autorange lower limit; -0.021 to 0.021
(amps).
2.1e-11
:LLIMit?
Query lower limit for autorange.
:AUTO?
Query state of autorange.
√
:DAMPing <b>
Enable or disable current damping.
OFF
Sec 4
:DAMPing?
Query state of damping
:RESistance
Path to configure ohms function:
√
:NPLCycles <NRf>
Set integration rate in line cycles (PLC); 0.01 to
10.
6 (60Hz)
5 (50Hz)
Sec 6
√
:NPLCycles?
Query NPLC.
√
:RANGe
Configure measurement range:
Sec 6
√
[:UPPer] <NRf>
Select range; 0 to 2.1e11 (ohms).
2.1e5
√
[:UPPer]?
Query range value.
√
:AUTO <b>
Enable or disable autorange.
(see Note)
√
:ULIMit <NRf>
Select autorange upper limit; -2.1e11 to
2.1e11 (ohms).
2.1e11
:ULIMit?
Query upper limit for autorange.
:LLIMit <NRf>
Select autorange lower limit; -2.1e11 to
2.1e11 (ohms).
2.1e3
:LLIMit?
Query lower limit for autorange.
:AUTO?
Query state of autorange.
√
:GUARd <b>
Enable or disable driven guard.
OFF
Sec 3
:GUARd?
Query state of driven guard.
:CHARge
Path to configure coulombs function:
√
:NPLCycles <NRf>
Set integration rate in line cycles (PLC); 0.01 to
10.
6 (60Hz)
5 (50Hz)
Sec 6
√
:NPLCycles?
Query NPLC.
√
:RANGe
Configure measurement range:
Sec 6
√
[:UPPer] <NRf>
Select range; -21e-6 to 21e-6 (coulombs).
2.1e-6
√
[:UPPer]?
Query range value.
√
:AUTO <b>
Enable or disable autorange.
(see Note)
√
Table 17-4 (cont.)
SENSe command summary
Summary of Contents for 6514
Page 203: ...14 CommonCommands...
Page 207: ...15 SCPISignalOriented MeasurementCommands...
Page 235: ...18 PerformanceVerification...
Page 254: ...19 Calibration...
Page 274: ...20 RoutineMaintenance...
Page 278: ...A Specifications...
Page 282: ...B StatusandErrorMessages...
Page 288: ...C GeneralMeasurement Considerations...
Page 294: ...D DDCEmulationCommands...
Page 299: ...E ExamplePrograms...
Page 306: ...F IEEE 488BusOverview...
Page 320: ...G IEEE 488andSCPI ConformanceInformation...
Page 324: ...H CalibrationOptions...