![background image](http://html2.mh-extra.com/html/keithley/6514/6514_instruction-manual_4064199083.webp)
Amps Measurements
4-11
source resistance values for various measurement ranges. Note that the recommended source
resistance varies by measurement range because the R
F
value also depends on the measurement
range.
Table 4-2
Minimum recommended source resistance values
Range
Minimum Recommended
Source Resistance
pA
1G
Ω
to 100G
Ω
nA
1M
Ω
to 100M
Ω
µA
1k
Ω
to 100k
Ω
mA
1
Ω
to 100
Ω
Current Source
R
F
C
F
-
+
V
noise
R
S
C
S
V
S
V
O
Z
F
Z
S
Model 6514 Ammeter
Figure 4-5
Source resistance
and capacitance