10-10
Limit Tests
Front panel operation
Limit test configuration
Most aspects of limit testing are configured from the limit configuration menu. When using
a component handler, the “start of test” (/STest, Test or Immediate) option is set from the arm
layer configuration menu.
Once in a menu structure, use the
and
keys to display menu items. Use the cursor
and
and the
and
keys to key in values. To change polarity of a value, place the cursor on “+”
or “-” and press
and
. To change range for the value, place the cursor on the range symbol
and use the
and
keys (P = pico,
N
= nano, µ = micro, m = milli, ^ = x1, K = kilo, M = mega,
G = giga, T = tera). A menu item or value is selected by pressing ENTER.
Digital output bit patterns
— An output pattern is set by selecting a decimal value (0 to 15)
that corresponds to the 4-bit BCD pattern of the output. To determine the output pattern value,
add up the decimal weight values for the desired HI lines:
For example, to set an output pattern to 0101 (lines 3 and 1 HI), set the output value to 5
(4 +1).
Limits configuration menu
The configuration menu for limits is structured as follows. Bullets denote the main items of
the menu. To access the menu, press SHIFT and then CONF-LIM.
•
LIMIT 1
— Configure limit 1 test:
CONTROL
— Enable or disable limit 1 test.
HILIM
— Set the HI limit (-9.999999T to +9.999999T), and set the “fail” digital out-
put bit pattern (0 to 15).
LOLIM
— Set the LO limit (-9.999999T to +9.999999T), and set the “fail” digital
output bit pattern (0 to 15).
•
LIMIT 2
— Configure limit 2 test:
CONTROL
— Enable or disable limit 2 test.
HILIM
— Set the HI limit (-9.999999T to +9.999999T), and set the “fail” digital out-
put bit pattern (0 to 15).
LOLIM
— Set the LO limit (-9.999999T to +9.999999T), and set the “fail” digital
output bit pattern (0 to 15).
•
PASS
— Set the digital output bit pattern for the “all tests pass” condition (0 to 15).
•
DIG CLR
(Digital Clear):
AUTO CLR
— Enable or disable auto-clear for the digital output.
DEL
(Delay) — Set the pass/fail pattern pulse width (0 to 60 sec with 10µsec resolu-
tion).
DIGOUT
— Set the digital output clear pattern (0 to 15).
Output HI Line:
Out 4
Out 3
Out 2
Out 1
Decimal Weight:
8
4
2
1
Summary of Contents for 6514
Page 203: ...14 CommonCommands...
Page 207: ...15 SCPISignalOriented MeasurementCommands...
Page 235: ...18 PerformanceVerification...
Page 254: ...19 Calibration...
Page 274: ...20 RoutineMaintenance...
Page 278: ...A Specifications...
Page 282: ...B StatusandErrorMessages...
Page 288: ...C GeneralMeasurement Considerations...
Page 294: ...D DDCEmulationCommands...
Page 299: ...E ExamplePrograms...
Page 306: ...F IEEE 488BusOverview...
Page 320: ...G IEEE 488andSCPI ConformanceInformation...
Page 324: ...H CalibrationOptions...