Calibration
19-15
10.
Press ENTER. Model 6514 will prompt for the negative full-scale charge value:
-20.0000nC
11.
Set the calibrator output voltage to -20.00000V. Calculate the actual charge from the cal-
ibrator voltage and actual standard capacitor value: Q = CV. Adjust Model 6514 display
to agree with the calculated charge value, then press ENTER to complete calibration of
the present range.
12.
Press EXIT to return to normal display.
13.
Repeat steps 4 through 12 for the 200nC through 20
µ
C ranges using Table 19-6 as a
guide.
Table 19-6
Coulombs calibration summary
Model 6514 range
Calibrator voltage
Standard capacitance
1
Calibration charge
2
20nC
+20.00000V
1nF
+20nC
-20.00000V
1nF
-20nC
200nC
+200.0000V
1nF
+200nC
-200.0000V
1nF
-200nC
2µC
+20.00000V
100nF
+2µC
-20.00000V
100nF
-2µC
20
µ
C
+200.0000V
100nF
+20µ
C
-200.0000V
100nF
-20
µ
C
1
Nominal value.
2
Based on nominal capacitance values. Calculate actual charge from calibrator voltage and actual standard
capacitance value: Q = CV. Calibrate positive full-scale, and negative full-scale values for each range.
Summary of Contents for 6514
Page 203: ...14 CommonCommands...
Page 207: ...15 SCPISignalOriented MeasurementCommands...
Page 235: ...18 PerformanceVerification...
Page 254: ...19 Calibration...
Page 274: ...20 RoutineMaintenance...
Page 278: ...A Specifications...
Page 282: ...B StatusandErrorMessages...
Page 288: ...C GeneralMeasurement Considerations...
Page 294: ...D DDCEmulationCommands...
Page 299: ...E ExamplePrograms...
Page 306: ...F IEEE 488BusOverview...
Page 320: ...G IEEE 488andSCPI ConformanceInformation...
Page 324: ...H CalibrationOptions...