18-4
Performance Verification
Recommended test equipment
Table 18-1 summarizes recommended verification equipment. You can use alternate equip-
ment, but keep in mind that test equipment accuracy will add to the uncertainty of each measure-
ment. Generally, the test equipment should have accuracy or uncertainty at least four times better
than corresponding Model 6514 specifications.
Table 18-1
Recommended verification equipment
Description
Manufacturer/model
Specifications
Calibrator
Fluke 5700A
DC voltage:
1
2V: 7ppm
20V: 5ppm
200V: 7ppm
DC current:
1
20
µ
A: 550ppm
200
µ
A: 100ppm
2mA: 55ppm
20mA: 55ppm
Resistance calibrator
Fluke 5450A
Nominal resistance:
1
1.9k
Ω
: 8ppm
19k
Ω
: 7.5ppm
190k
Ω
: 8.5ppm
1.9M
Ω
: 11.5ppm
19M
Ω
: 30ppm
100M
Ω
: 120ppm
Electrometer calibration
Keithley Model 5156
Nominal resistance:
2
standard
100M
Ω
: 200ppm
1G
Ω
: 300ppm
10G
Ω
: 400ppm
100G
Ω
: 500ppm
Nominal capacitance:
2
1nF: 1,000ppm
100nF: 1,000ppm
Notes:
1. 90-day, 23˚ ±5˚C full-range accuracy specifications shown,
2. 23˚ ±3˚C accuracy of characterization.
3. Connect red and black clips to make triax short.
Summary of Contents for 6514
Page 203: ...14 CommonCommands...
Page 207: ...15 SCPISignalOriented MeasurementCommands...
Page 235: ...18 PerformanceVerification...
Page 254: ...19 Calibration...
Page 274: ...20 RoutineMaintenance...
Page 278: ...A Specifications...
Page 282: ...B StatusandErrorMessages...
Page 288: ...C GeneralMeasurement Considerations...
Page 294: ...D DDCEmulationCommands...
Page 299: ...E ExamplePrograms...
Page 306: ...F IEEE 488BusOverview...
Page 320: ...G IEEE 488andSCPI ConformanceInformation...
Page 324: ...H CalibrationOptions...