Range, Units, Digits, Rate, and Filters
6-11
Programming example
The following command sequence configures and enables both filters:
‘ Median Filter:
MED:RANK 5
‘ Set rank to 5.
MED ON
‘ Enable median filter.
‘ Digital Filter:
AVER:COUN 20
‘ Set filter count to 20.
AVER:TCON MOV
‘ Select moving filter.
AVER ON
‘ Enable digital filter.