Coulombs Measurements
5-5
SCPI programming
A) SENSe:FUNCtion <name>
Parameters
‘CHARge’
Coulombs function
‘CURRent’
Amps function
‘VOLTage’
Volts function
‘RESistance’
Ohms function
Note that the parameter names are enclosed in single quotes (‘). However, double quotes (“)
can instead be used. Each measurement function “remembers” its own unique range setting.
B) SENSe:DATA?
This command does not trigger a reading. It simply returns the last “raw” reading string. It
will not return the result of any instrument calculation. The reading reflects what is applied to
the input.
To return a fresh (new) reading, you can send the INITiate command to trigger one or more
readings before sending :DATA?. Details on INITiate are provided in Section 9.
While Model 6514 is busy performing measurements, the :DATA? command will not return
the reading string until the instrument finishes and goes into the idle state.
Table 5-1
SCPI commands — coulombs function
Commands
Description
Default
Ref
[SENSe]
SENSe Subystem:
:FUNCtion ‘CHARge’
Select coulombs function.
VOLT
A
:DATA?
Return latest “raw” reading.
B
:CHARge
:ADIScharge
Auto discharge:
[:STATe] <b>
Enable or disable auto discharge.
OFF
:LEVel <NRf>
Set auto discharge level; -2.1e-5 to 2.1e-5.
2e-6
INITiate
Trigger one or more readings.
B
READ?
Trigger and return reading(s).
B
Summary of Contents for 6514
Page 203: ...14 CommonCommands...
Page 207: ...15 SCPISignalOriented MeasurementCommands...
Page 235: ...18 PerformanceVerification...
Page 254: ...19 Calibration...
Page 274: ...20 RoutineMaintenance...
Page 278: ...A Specifications...
Page 282: ...B StatusandErrorMessages...
Page 288: ...C GeneralMeasurement Considerations...
Page 294: ...D DDCEmulationCommands...
Page 299: ...E ExamplePrograms...
Page 306: ...F IEEE 488BusOverview...
Page 320: ...G IEEE 488andSCPI ConformanceInformation...
Page 324: ...H CalibrationOptions...