![background image](http://html2.mh-extra.com/html/keithley/6514/6514_instruction-manual_4064199035.webp)
Getting Started
1-13
Remote operation
Saving and restoring user setups
— The *SAV and *RCL commands are used to save and
recall user setups. These commands are documented in Section 14.
Restoring factory or GPIB default setups
— The SYSTem:PRESet command returns
Model 6514 to the factory defaults and the *RST command returns it to the GPIB defaults. The
*RST command is documented in Section 14 and SYSTem:PRESet is covered in Section 16
(SYSTem Subsystem).
Selecting power-on setup
— The SYSTem:POSetup command is used to select which setup
to return to on power-up. The SYSTem:POSetup command is documented in Section 16
(SYSTem Subsystem).
Table 1-2
Default settings
Setting
Factory
GPIB
Arm Layer (CONF-ARM):
Arm-In Source Event
IMM
IMM
Arm Count
INF
1
Input Trigger Link Line
1
1
Source Bypass
NEVER
NEVER
Output Trigger Link Line
2
2
Output Trigger
Off
Off
Auto Discharge
Off
Off
Level
2e-6
2e-6
Buffer (STORE):
Disabled
Disabled
Count
No effect
No effect
Digital Filter (AVG):
Off
Off
Count
10
10
Type
Moving
Moving
Display Resolution (DIGIT)
5
½
-digits
5
½
-digits
Function
Volts
Volts
Guard
Off
Off
GPIB:
No effect (On at factory)
No effect (On at factory)
Address
No effect (14 at factory)
No effect (14 at factory)
Language
No effect (SCPI at factory) No effect (SCPI at factory)
Summary of Contents for 6514
Page 203: ...14 CommonCommands...
Page 207: ...15 SCPISignalOriented MeasurementCommands...
Page 235: ...18 PerformanceVerification...
Page 254: ...19 Calibration...
Page 274: ...20 RoutineMaintenance...
Page 278: ...A Specifications...
Page 282: ...B StatusandErrorMessages...
Page 288: ...C GeneralMeasurement Considerations...
Page 294: ...D DDCEmulationCommands...
Page 299: ...E ExamplePrograms...
Page 306: ...F IEEE 488BusOverview...
Page 320: ...G IEEE 488andSCPI ConformanceInformation...
Page 324: ...H CalibrationOptions...