DISPlay, FORMat, and SYSTem
16-9
A) SYSTem:PRESet
Returns the instrument to states optimized for front panel operation. SYSTem:PRESet
defaults are listed in the SCPI tables in Section 17.
B) SYSTem:TIME:RESet
Resets the absolute timestamp to 0 seconds. The timestamp also resets when power is cycled,
or after the instrument is on for 99,999.99 seconds. The TRACe:TSTamp:FORMat command is
used to select the absolute timestamp. See Section 8 (Buffer) for details.
C) SYSTem:POSetup <name>
Parameters
RST
Power-up to *RST defaults
PRESet
Power-up to SYSTem:PRESet defaults
SAVx
Power-up to setup stored in memory (x = memory location 0,
1, 2, 3 or 4)
The *RST and SYSTem:DEFaults are listed in the SCPI tables in the Section 17. A setup is
saved in memory using the *SAV command. See Section 14 (Common Commands) for details.
D) SYSTem:VERSion
Read the version of the SCPI standard being used by Model 6514. Example response mes-
sage: 1996.0.
E) SYSTem:KEY <NRf>
Parameters
1
=
SHIFT key
17 =
LOCAL key
2
=
V key
18 =
AVG key
3
=
I key
19 =
MEDIAN key
4 =
Ω
key
20 =
REL key
5
=
Q key
21 =
LIMIT key
6
=
XFDBK key
22 =
DIGIT key
7
=
ZCHECK key
23 =
RATE key
8
=
ZCOR key
24 =
Cursor left arrow key
9 =
-------
25 =
-------
10 =
-------
26 =
STORE key
11 =
RANGE up arrow key
27 =
RECALL key
12 =
AUTO key
28 =
DELAY key
13 =
RANGE down arrow key
29 =
DAMP key
14 =
ENTER key
30 =
HALT key
15 =
Cursor right arrow key
31 =
TRIG key
16 =
GRD key
32 =
EXIT key
Summary of Contents for 6514
Page 203: ...14 CommonCommands...
Page 207: ...15 SCPISignalOriented MeasurementCommands...
Page 235: ...18 PerformanceVerification...
Page 254: ...19 Calibration...
Page 274: ...20 RoutineMaintenance...
Page 278: ...A Specifications...
Page 282: ...B StatusandErrorMessages...
Page 288: ...C GeneralMeasurement Considerations...
Page 294: ...D DDCEmulationCommands...
Page 299: ...E ExamplePrograms...
Page 306: ...F IEEE 488BusOverview...
Page 320: ...G IEEE 488andSCPI ConformanceInformation...
Page 324: ...H CalibrationOptions...