20-4
Routine Maintenance
Front panel tests
The front panel tests are summarized in Table 20-2. To run a test, simply press SHIFT then
TEST, then scroll through the menu choices and press ENTER.
DISP test
The display test allows you to verify that each segment and annunciator in the vacuum fluo-
rescent display is working properly. Perform the following steps to run the display test:
1.
Press SHIFT and then TEST to access the self-test options.
2.
Use the up or down RANGE key to display "TEST: DISP".
3.
Press ENTER to start the test. There are four parts to the display test. Each time ENTER
is pressed, the next part of the test sequence is selected. The four parts of the test
sequence are as follows:
•
All annunciators are displayed.
•
The segments of each digit are sequentially displayed.
•
The 12 digits (and annunciators) are sequentially displayed.
•
The annunciators located at either end of the display are sequentially displayed.
4.
When finished, abort the display test by pressing EXIT. The instrument returns to normal
operation.
KEY test
The KEY test allows you to check the functionality of each front panel key. Perform the fol-
lowing steps to run the KEY test:
1.
Press SHIFT and then TEST to access the self-test options.
2.
Use the up or down RANGE key to display "TEST: KEY".
3.
Press ENTER to start the test. When a key is pressed, the label name for that key is dis-
played to indicate that it is functioning properly. When the key is released, the message
"NO KEY PRESS" is displayed.
4.
Pressing EXIT tests the EXIT key. However, the second consecutive press of EXIT
aborts the test and returns the instrument to normal operation.
Table 20-2
Front panel tests
Test
Description
DISP
Test display
KEY
Test front panel keys
Summary of Contents for 6514
Page 203: ...14 CommonCommands...
Page 207: ...15 SCPISignalOriented MeasurementCommands...
Page 235: ...18 PerformanceVerification...
Page 254: ...19 Calibration...
Page 274: ...20 RoutineMaintenance...
Page 278: ...A Specifications...
Page 282: ...B StatusandErrorMessages...
Page 288: ...C GeneralMeasurement Considerations...
Page 294: ...D DDCEmulationCommands...
Page 299: ...E ExamplePrograms...
Page 306: ...F IEEE 488BusOverview...
Page 320: ...G IEEE 488andSCPI ConformanceInformation...
Page 324: ...H CalibrationOptions...