Twister / Twister Junior
Service Manual
Issued: 21.01.02
Page 185
Check Remedy
•
Check electrical functions
on control!
•
BUS plug defective?
•
BUS cable damaged?
•
Drive lever tight or
damaged?
•
Screw connections loose?
•
Parts / Support defective?
•
Motor defective?
•
Switchbox damaged?
⇒
See chapter 18.2.
⇒
Replace BUS cable.
⇒
Replace BUS cable.
⇒
Replace control.
⇒
Tighten screws.
⇒
Replace parts / support.
⇒
Replace motor.
⇒
Replace switchbox.
Note
:
•
On the motor, a continuity
test can be performed.
•
Also, a performance test
can be performed by
applying a voltage (24 V).
•
The tests must be
performed in the same
way as described in chapt.
17for the actuator motors.
19.5.1 Disassembly of control
•
Detach BUS cable (26) from the control (1).
•
Loosen screw (30), open clamp (29), and detach control (1) from its attachment.
19.5.2 Reassembling the control
•
Reassembly takes place in the reverse order.
19.5.3 Disassembly of the motor
•
Detach motor connection cable (38) from wheelchair and lay bare.
•
Remove securing ring (17) from bolt (14) so that the motor (8) can be freed from the back
support rod (2).
•
Loosen nut (37) completely and detach together with washer (25).
•
Support motor (8), remove screw (19) and detach motor from swivel arm (3).
19.5.4 Reassembly of the motor
•
Reassembly takes place in the reverse order.